Electrochemical impedance spectroscopy
Jump to navigation
Jump to search
Electrochemical Impedance Spectroscopy
- This SOP discusses how to conduct an EIS test on a sample with an isolated, known surface area. This process was originally devised to test the heads of screws with various coating systems.
Safety
- Eye protection should be worn when handling glass.
- Appropriate gloves, chemical PPE should be worn when handling liquids.
- If you using a copper sulfate solution or any other irritant, take care to not spill or expose one's self to the liquid.
- Avoid contact with liquids that have electricity running through them, do not put fingers within the vessel.
Procedure
- Prepare samples for EIS Testing.
- Add desired amount of the solution chosen to an electrochemical cell glassware
- Add a magnetic stirrer
- Connect the sample to a section of wire.
- Lower the sample into the electrochemical cell glassware.
- Position the sample so that the desired area is submerged in the solution.
- Add the counter electrodes to the glassware
- add the luggin probe
- Position the luggin probe close to the sample surface that is being tested
- Turn the power supply on
- Connect the power supply wires in the following order
- WE to the sample wire
- CE to counter electrode
- RE to reference electrode
- SE to the sample wire
- Run an open circuit potential test, then potentiostatic test, then EIS test using the software.
- The potentiodynamic test should be -0.5V to 0.5V compared to open circuit
Sample Preparation
- All Samples must have certain requirements to be compatible for EIS testing.
- Sample must be a conductive material.
- Sample must have an electrical connection to the power supply.
- Sample must have all non-conductive material removed at contact point to wire.